Auxiliary functions – Hanna Instruments HI 903 User Manual
Page 111
Advertising

7-5
AUXILIARY FUNCTIONS
To prime the burette, select
Prime Burette
from the
Burette
screen. Enter the number of
rinses and press
.
The number of burette rinses can be set between 1 and 5 (we recommend at least three
rinses to assure that the air bubbles are completely removed).
7.2.2
Rinse Tip
A 0.25 mL dose of titrant will be dispensed from the burette when this operation is selected.
This operation will eliminate any contamination from the anti-diffusion dispensing tip.
7.2.3
Manual Dispense
Manual Dispense
allows a defined titrant volume to be be dosed. Select the
Manual Dispense
option and press
. The
Manual Volume Dispense
screen will become active and the
display will prompt you to enter the desired volume to be dispensed.