• After the pump calibration is complete a confirmation message is displayed for a few seconds,

then the instrument will return to the titrator calibration menu and will display the new time
and date for the last pump calibration.

Notes: • The calibration of the pump is independent of the selected range and concentration unit.

• If an erroneous situation is encountered during the calibration, an error message is

displayed and the calibration can be restarted by pressing Restart.

• If the calibration doesn’t complete within 6 minutes the error message “Too much

standard” will be displayed and the calibration can be restarted by pressing Restart
after a new sample is prepared.