Accessories, Battery replacement – Hanna Instruments HI 95747C User Manual

Page 18

Advertising
background image

18

REAGENT SETS
HI 95747-01 Reagents for 100 tests
HI 95747-03 Reagents for 300 tests
OTHER ACCESSORIES

HI 95747-11

CAL CHECK

TM

Standard Cuvets (1 set)

HI 710009

Blue rubber boot

HI 710010

Orange rubber boot

HI 721310

9V battery (10 pcs)

HI 731318

Tissue for wiping cuvets (4 pcs)

HI 731331

Glass cuvets (4 pcs)

HI 731335

Caps for cuvets (4 pcs)

HI 93703-50 Cuvets cleaning solution (230 mL)

ACCESSORIES

Battery replacement must only take place in a
non-hazardous area.
Simply slide off the battery cover on the back of
the meter. Detach the battery from the terminals
and attach a fresh 9V battery while paying attention
to the correct polarity. Insert the battery and
replace the cover.

BATTERY REPLACEMENT

FACTORY CALIBRATION RESET
It is possible to restore factory calibration:
• Enter the

calibration mode

by holding CAL

CHECK

TM

for three seconds

.

• Hold ZERO for 10 seconds. The display will

show for 2 seconds “F.CAL” and "747" appears.
The factory calibration is automatically restored
and the instrument is ready for measurement.