Hanna Instruments HI 9811 User Manual
Page 11

11
The TDS value in aqueous solutions is directly proportional to the
conductivity. The ratio between the two parameters depends on the
solution.
HI 9811-5 model has a fixed conversion factor set to 0.5. This
means that 1
µ
S/cm is equal to 0.5 mg/L of TDS.
Accessories needed:
• Use
HI 70031 (1413
µ
S/cm) EC calibration solution or
HI 70032 (1382 ppm, or mg/L) TDS calibration solution.
Note: The conversion between EC and TDS is made by a built-in
circuit, hence it is requested to calibrate the meter only in
EC or TDS range. The other range is thus automatically
calibrated.
PROCEDURE
• Pour approximately 4 cm (1½") of a conductivity calibration
solution (e.g.
HI 70031) into a beaker. If possible, use plastic
beaker to minimize any EMC interference.
• Immerse the probe in the solution.
• Wait for a couple of minutes for thermal equilibrium to be
reached.
• Tap the probe on the bottom, then shake it lightly while rotating
to make sure no air bubbles remain trapped inside the probe.
• Press the
µ
S/cm (or mg/L) key.
• Turn the EC/TDS calibration knob until
the display shows the EC or TDS reading
at 25°C.