Hanna Instruments HI 96715 User Manual

Page 8

Advertising
background image

14

15

• Press ZERO/CFM and the lamp, cuvette

and detector icons will appear on the
display, depending on the measurement
phase.

• After a few seconds, the display will show

“-0.0-”

. The meter is now zeroed and

ready for measurement.

• Remove the cuvette, open it and add 4

drops of HI 93715A-0 First Reagent (6
drops in case of seawater analysis). Swirl
the cuvette gently to mix.

• Add 4 drops of HI 93715B-0 Second

Reagent to the cuvette (10 drops in case
of seawater analysis). Replace the cap
and swirl the cuvette to mix.

• Replace the cuvette into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.

• Press and hold READ/

READ/

READ/

READ/

READ/ /TIMER

/TIMER

/TIMER

/TIMER

/TIMER

for three

seconds.

The display will show the

countdown prior to measurement. The
beeper is playing a beep at the end of
countdown period.

Alternatively, wait for 3 minutes and 30
seconds and just press READ/

READ/

READ/

READ/

READ/ /TIMER

/TIMER

/TIMER

/TIMER

/TIMER.

In both cases, the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.

• At the end of measurement, the instrument

directly displays concentration in mg/L of
ammonia nitrogen (NH

3

-N) on the LCD.

• To convert the reading to mg/L of

ammonia (NH

3

), multiply by a factor of

1.216.

VALIDATION PROCEDURE

Use the validation procedure to ensure that the instrument is properly calibrated.
Warning:

Do not validate the instrument with any standard solutions other than the HANNA CAL

CHECK

TM

Standards, otherwise erroneous results will be obtained. For accurate validation, please

perform test at room temperature: 18 to 25°C (64.5 to 77.0°F)

Turn the meter on by pressing ON/OFF.

• When the beeper sounds briefly and the

LCD displays dashes, the meter is ready.

INTERFERENCES

Organic compounds like: chloramines, various aliphatic and aromatic amines, glycine,
(positive error).
To eliminate these interferences distillation is required.

Organic compounds like: aldehydes, alcohols (e.g. ethanol) or acetone (negative error).
To eliminate these interferences distillation is required.

Sulfide (S

2-

): may cause turbidity.

Hardness above 1 g/L as Calcium Carbonate (CaCO

3

).

Advertising