Hanna Instruments HI 96701C User Manual

Page 14

Advertising
background image

14

• Remove the cuvette.

• Add the content of one packet of

HI 93701-0

Free Chlorine reagent.

Replace the cap and shake gently for
20 seconds (or 2 minutes in case of
seawater analysis).

• Replace the cuvette into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.

• Press and hold READ/ /TIMER

for three

seconds.

The display will show the

countdown prior to measurement. The
beeper is playing a beep at the end of
countdown period.
Alternatively, wait for 1 minute and just
press READ/ /TIMER.
In both cases, the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.

• At the end of measurement, the

instrument directly displays concentration
in mg/L of free chlorine on the LCD.

INTERFERENCES
Bromine (positive error)
Ozone (positive error)
Iodine (positive error)