Hanna Instruments HI 95729 User Manual

Page 8

Advertising
background image

8

KEYPAD DESCRIPTION
• ON/OFF: to turn the meter on and off.
• ZERO: to zero the meter prior to measurement.
• READ/TIMER: this is a bi-functional key, just press for making a measurement, or hold the

key for 3 seconds to start a pre-programmed countdown prior to measurement.

• CAL CHECK: this bi-functional key allows to perform validation of the instrument just pressing

the CAL CHECK key, or hold the key for 3 seconds to enter the

calibration mode.

OPERATING MODES

Measurement mode: default operation mode, enables both validation and measurement.

Calibration mode: may be entered by keeping CAL CHECK pressed

for three seconds (

the “CAL”

icon appears), it enables

calibration of the instrument.

DISPLAY ELEMENTS DESCRIPTION

1) four digit main display area.
2) CAL: appears when the instrument operates in the

calibration mode.

3) the Month and Day icons appear above the number of month and day when a date is

displayed.

4) the blinking “Battery” icon appears when the battery voltage is getting low.

This prompt appears for 1 second each time the instrument is
turned on.

This indicates that the meter is in a ready state and zeroing can
be performed.
Sampling in Progress. This

flashing prompt

appears each time the

meter is performing a measurement.

GUIDE TO DISPLAY CODES

13

b

l

a

n

k

• Replace the cap to the cuvet, and shake

gently a couple of times.

• Place the cuvet with the reacted deionized

water (blank) into the holder and ensure that
the notch on the cap is positioned securely
into the groove.

• Hold READ/TIMER

for three seconds

and the

display will show the countdown or,
alternatively, wait for two minutes.

• Press ZERO and "SIP" will blink on the

display.
Wait for a few seconds and the display will
show "-0.0-". Now the meter is zeroed and
ready for measurement.

• Remove the blank from the instrument.
• Remove the cap from the cuvet and discard

the blank.

• Rinse the cuvet a couple of times with the

unreacted sample and discard it.
Turn upside down the empty cuvet and shake
it to remove excess of liquid.

• Fill the same cuvet with 10 mL of unreacted

sample.
For most accurate results, use of another
clean class A laboratory pipette is strongly
recommended.
Alternatively, fill the cuvet up to the 10 mL
mark.

• Add 2 mL of HI 93729-0 Fluoride Reagent

to the cuvet by use of the 2000 µL automatic
pipette.
For a correct use of the automatic pipette,
please follow the related Instruction Sheet.

2

m

L

Advertising
This manual is related to the following products: