INTERFERENCES

• Positive error:

- Bromine
- Chlorine dioxide
- Iodine
- Ozone
- Oxidized Manganese and Chromium

• Alkalinity above 250 mg/L CaCO

3

or acidity above 150 mg/L CaCO

3

will not reliably develop

the full amount of color or it may rapidly fade.
To resolve this, neutralize the sample with diluted HCl or NaOH.

• In case of water with hardness greater than 500 mg/L CaCO

3

, shake the sample for

approximately 2 minutes after adding the powder reagent.

Chlorine

•

For free chlorine measurement: add 3 drops
of HI 93701A-T DPD1 indicator and 3
drops of HI 93701B-T DPD1 buffer to an
empty cuvet.
Immediately add 10 mL of unreacted
sample. Replace the cap and shake gently.

•

Immediately reinsert the cuvet into the
holder and ensure that the notch on the
cap is positioned securely into the groove.

• Press READ and “SIP” will blink during

measurement.

• The instrument directly displays

concentration in mg/L of free chlorine on
the Liquid Crystal Display.

• Then,

for total chlorine measurement, remove

the cuvet and add to the reacted sample 3
drops of HI 93701C-T DPD3 solution.
Replace the cap and swirl gently.

• Reinsert the cuvet into the holder and

ensure that the notch on the cap is
positioned securely into the groove.

• Wait for 2’ and 30” then press READ and

“SIP” will blink during measurement.

• The instrument directly displays

concentration in mg/L of total chlorine on
the Liquid Crystal Display.

• Remove the cuvet.

Chlorine