Hanna Instruments HI 95721 User Manual

Page 9

Advertising
background image

9

This prompt appears for 1 second each time the instrument is turned
on.

This indicates that the meter is in a ready state and zeroing can be
performed.

Sampling in Progress. This flashing prompt appears each time the
meter is performing a measurement.

This indicates that the meter is in a zeroed state and measurement
can be performed.

Light over range: the cuvet is not inserted correctly and an excess

ambient light is reaching the detector. If the cover is properly installed,
then contact your dealer or the nearest Hanna Customer Service
Center.

The blinking "BAT" indicates that the battery voltage is getting low
and the battery needs to be replaced.

This indicates that the battery is dead and must be replaced. Once
this indication is displayed, the meter will lock up. Change the

battery and restart.

“Configuration”: this indicates that the meter has lost its configuration.
Contact your dealer or the nearest Hanna Customer Service Center.

CALIBRATION MODE MESSAGES

The date of the last calibration appears on the LCD each time the
calibration mode

is entered. If calibration is performed for the first

time, “F.CAL” appears.

This indicates that the factory calibration is selected.

GUIDE TO DISPLAY CODES

12

• It is important that the sample does not contain any

debris. This would corrupt the reading.

• Each time the cuvet is used, the cap must be

tightened to the same degree.

• Whenever the cuvet is placed into the measurement

cell, it must be dry outside, and completely free of
fingerprints, oil or dirt. Wipe it thoroughly with
HI 731318 (tissue for wiping cuvets, see chapter
ACCESSORIES) or a lint-free cloth prior to insertion.

• Shaking the cuvet can generate bubbles in the

sample, causing higher readings. To obtain accurate
measurements, remove such bubbles by swirling or
by gently tapping the vial.

• Do not let the reacted sample stand too long after

reagent is added, or accuracy will be lost.

• It is possible to take multiple readings in a row, but

it is recommended to take a new zero reading for
each sample and to use the same cuvet for zeroing
and measurement.

• After the reading it is important to discard

immediately the sample, otherwise the glass might

become permanently stained.

• All the reaction times reported in this manual are

referred to 20°C (68°F). As a general rule of thumb,
they should be doubled at 10°C (50°F) and halved
at 30°C (86°F).

• In order to maximize accuracy, prior to a

measurement follow the validation procedure to
be sure that the instrument is properly calibrated. If

necessary, calibrate the instrument.

Advertising
This manual is related to the following products: