Hanna Instruments HI 95707 User Manual
Page 5

5
16
Range
0.000 to 0.600 mg/L (as NO
2
–
-N)
Resolution
0.001 mg/L
Precision
±0.001 mg/L @ 0.100 mg/L
Typical EMC Deviation
±0.001 mg/L
Light Source
Tungsten lamp with narrow band interference filter @ 525 nm
Light Detector
Silicon Photocell
Method
Adaptation of an EPA approved method. The reaction between Nitrite
and the reagent causes a pink tint in the sample.
Environment
0 to 50°C (32 to 122°F); max 95% RH non-condensing
Battery Type
1 x 9 volt
Auto-Shut off
After 10' of non-use in measurement mode;
after 1 hour of non-use in calibration mode.
Dimensions
180 x 83 x 46 mm (7.1 x 3.3 x 1.8")
Weight
290 g (10 oz.)
REQUIRED REAGENTS
Code
Description
Quantity/test
HI 93707-0
Nitrite Low Range Reagent
1 packet
SPECIFICATIONS
Precision is how closely repeated
measurements agree with each other.
Precision is usually expressed as standard
deviation (SD).
Accuracy is defined as the nearness of a test
result to the true value.
Although good precision suggests good
accuracy, precise results can be inaccurate. The
figure explains these definitions.
In a laboratory using a standard solution of
0.100 mg/L Nitrogen-Nitrite and a
representative lot of reagent, an operator
obtained with a single instrument a standard
deviation of 0.001 mg/L.
SPECIFICATIONS
PRECISION AND ACCURACY
CALIBRATION PROCEDURE
• Turn the meter on by pressing ON/OFF.
When the LCD displays “707”, it is ready.
• Enter the calibration mode by holding CAL CHECK
for three seconds.
• The date of the last calibration appears (e.g.:
month “01”, day “08”). “F.CAL” means that the
factory calibration is selected.
Note: at this point it is possible to reset the
instrument to restore FACTORY CALIBRATION, see
instructions page 18.
• Place the CAL CHECK™ Standard Cuvet A into
the holder and ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will blink on the display.
• After a few seconds the display will show
“-0.0-”. The meter is now zeroed and ready for
calibration.
• Remove the cuvet.
• Place the CAL CHECK™ Standard HI 95707-11
Cuvet B into the holder and ensure that the notch
on the cap is positioned securely into the groove.
Note: it is possible to interrupt calibration procedure at any time by pressing ON/OFF.
Warning: do not validate or calibrate the instrument with standard solutions other than the Hanna CAL
CHECK™ Standards, otherwise erroneous results will be obtained. For accurate Validation and Calibration
results, please perform test at room temperature: 18 to 25°C (64.5 to 77.0°F).
or
CALIBRATION PROCEDURE