Hanna Instruments HI 83741 User Manual

Page 5

Advertising
background image

5

Range

0.0-15.0 mg/L

Resolution

0.1 mg/L

Precision

SD ±0.4 mg/L @ 4.0 mg/L

Light Source

Tungsten lamp with narrow band interference filter @ 560 nm

Light Detector

Silicon Photocell

Method

The reaction between Iron and the reagents causes a purple tint in
the sample.

Environment

0 to 50°C (32 to 122°F); max 95% RH non-condensing

Battery Type

4 x 1,5 volt AA batteries / 12 to 20 VDC through voltage adapter

Auto-Shut off

After 15' of non-use in measurement mode.

Dimensions

225 x 85 x 80 mm (8.7 x 3.3 x 3.1")

Weight

500 g (17,6 oz.)

REQUIRED REAGENTS

Code

Description

Quantity/test

HI 83741A-0

Iron Reagent A

1 packet

HI 83741B-0

Iron Reagent B

1 packet

HI 83742-0

Wine solvent-1

9 mL

SPECIFICATIONS

Precision is how closely repeated measurements agree
with each other. Precision is usually expressed as
standard deviation (SD). Accuracy is defined as the
nearness of a test result to the true value.
Although good precision suggests good accuracy,
precise results can be inaccurate. The figure explains
these definitions.
In a laboratory using a standard solution of 4.0
mg/L iron and a representative lot of reagent, an
operator obtained with a single instrument a standard
deviation of

0.4 mg/L.

PRECISION AND ACCURACY

16

Battery replacement must only take place in a non-
hazardous area.
The blinking “ ” will appear when the batteries power
gets low.
When batteries are completely discharged, “0% bAtt”
will appear and after two seconds the instrument is
switched off.
Remove the battery cover from the bottom of the
instrument and change the old batteries with 4 fresh
1.5V batteries, paying attention to the correct polarity.
Replace the cover.

BATTERIES REPLACEMENT

REAGENT SETS
HI 83741-20

Iron reagents set for wine (20 tests)

OTHER ACCESSORIES
HI 740027P

1.5V AA batteries (10 pcs)

HI 731318

Tissue for wiping cuvets (4 pcs)

HI 731321

Glass cuvets (4 pcs)

HI 731325W

Caps for cuvets (4 pcs)

HI 93703-50

Cuvets cleaning solution (230 mL)

HI 731341

1000

µL automatic pipette

HI 731351

Plastic tips for 1000

µL automatic pipette (25 pcs)

ACCESSORIES

Advertising