Hanna Instruments HI 83225-01 User Manual
Page 48

48
• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the cuvette and add the content of one
packet of HI 93750B-0 reagent. Replace the cap
and gently mix for one minute by slowly turning the
cuvette upside down.
• Reinsert the cuvette into the instrument.
• Press Timer and the display will show the countdown prior to
the measurement or, alternatively, wait for 2 minutes and press
Read. When the timer ends the meter will perform the reading.
The instrument displays the results in mg/L (ppm) of
potassium (K).
• Press
or
to access the second level functions.
• Press the Chem Frm key to convert the result in mg/L of potassium oxide (K
2
O).
• Press
or
to return to the measurement screen.
INTERFERENCES
Interferences may be caused by:
Ammonium above 10 ppm
Calcium above 10000 ppm as CaCO
3
Chloride above 12000 ppm
Magnesium above 8000 ppm as CaCO
3
Sodium above 8000 ppm
Potassium LR