Cyanuric Acid
SPECIFICATIONS
Range
0 to 80 mg/L
Resolution
1 mg/L
Accuracy
±1 mg/L ±15% of reading at 25 °C
Typical EMC
±1 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter @ 525 nm
Method
Adaptation of the turbidimetric method. The reaction between cyanuric acid and the
reagent causes a white suspension in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93722-0
Powder reagent
1 packet
REAGENT SETS
HI 93722-01 Reagents for 100 tests
HI 93722-03 Reagents for 300 tests
For other accessories see page 66.
MEASUREMENT PROCEDURE
• Select the
Cyanuric Acid method using the procedure
described in the
Method Selection section (see page 12).
• Fill the cuvette with 10 mL of unreacted sample (up to the
mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for
measurement.
10 mL
CYANURIC ACID