Hanna Instruments HI 9812-5N User Manual

Page 10

Advertising
background image

10

Accessories needed:

• Use

HI 70031 (1413

µ

S/cm) EC calibration solution or

HI 70032 (1382 ppm) TDS calibration solution.

Note: The conversion between EC and TDS is made by a built-in

circuit, hence it is requested to calibrate the meter only in EC
or TDS range. The other range is thus automatically calibrated.

PROCEDURE
• Pour approximately 4 cm (1½") of a conductivity calibration

solution (e.g.

HI 70031) into a beaker. If possible, use plastic

beaker to minimize any EMC interference.

• Immerse the probe in the solution.
• Wait for a couple of minutes for thermal equilibrium to be

reached.

• Tap the probe on the bottom, then shake it lightly while rotating

to make sure no air bubbles remain trapped inside the probe.

• Press the

µ

S/cm (or ppm) key.

• Turn the EC/TDS calibration knob until

the display shows the EC or TDS reading
at 25°C.

E C / T D S

E C / T D S

E C / T D S

E C / T D S

E C / T D S C A L I B R A T I O N

C A L I B R A T I O N

C A L I B R A T I O N

C A L I B R A T I O N

C A L I B R A T I O N

Advertising