The setup mode is used to view or change the instrument parameters.

To enter setup press ALT+FNC when the
meter is in measurement mode.

"Set" is displayed on the upper LCD. The
lower LCD displays the blinking code of the
current setup item.

Select the desired setup item using
the

or

key.

Press ALT+CFM to confirm.

Note: If ALT+FNC are pressed before item confirmation, the meter

will escape and return to measurement mode.

Once the desired setup item has been
selected, its current value blinks (if it is a
changeable parameter).

To change the value use the

or

key.

Press ALT+CFM to confirm.

Note: Press ALT+FNC before confirmation to escape without chang-

ing the previously set value.

SETUP

The conductivity of an aqueous solution is the measure of its ability to
carry an electrical current by means of ionic motion.
The conductivity invariably increases with increasing temperature.
It is affected by the type and number of ions in the solution and by
the viscosity of the solution itself. Both parameters are temperature
dependent. The dependency of conductivity on temperature is ex-
pressed as a relative change per degree Celsius at a particular
temperature, commonly as percent per °C.
The following table lists the temperature dependence of the HANNA
instruments

®

calibration solutions.

°C

°F

HI 7030

HI 7031

HI 7033

HI 7034

HI 7035

HI 7039

HI 8030

HI 8031

HI 8033

HI 8034

HI 8035

HI 8039

(μS/cm)

(μS/cm)

(μS/cm)

(μS/cm)

(μS/cm)

(μS/cm)

0

32

7150

776

64

48300

65400

2760

5

41

8220

896

65

53500

74100

3180

10

50

9330

1020

67

59600

83200

3615

15

59

10480

1147

68

65400

92500

4063

16 60.8

10720

1173

70

67200

94400

4155

17 62.6

10950

1199

71

68500

96300

4245

18 64.4

11190

1225

73

69800

98200

4337

19 66.2

11430

1251

74

71300

100200

4429

20

68

11670

1278

76

72400

102100

4523

21 69.8

11910

1305

78

74000

104000

4617

22 71.6

12150

1332

79

75200

105900

4711

23 73.4

12390

1359

81

76500

107900

4805

24 75.2

12640

1386

82

78300

109800

4902

25

77

12880

1413

84

80000

111800

5000

26 78.8

13130

1440

86

81300

113800

5096

27 80.6

13370

1467

87

83000

115700

5190

28 82.4

13620

1494

89

84900

117700

5286

29 84.2

13870

1521

90

86300

119700

5383

30

86

14120

1548

92

88200

121800

5479

31 87.8

14370

1575

94

90000

123900

5575

CONDUCTIVITY VS TEMPERATURE CHART