Installing the ampbox – Kipp&Zonen AMPBOX Amplifier User Manual

Page 13

Advertising
background image

Before installing the AMPBOX you should be aware that we do not supply the cable between the AMPBOX and the data acquisition
system. In fact the AMPBOX is supplied without any cables. The cable from the radiometer to the AMPBOX can be ordered together
with the radiometer.

Note

Install the AMPBOX as close as possible to the radiometer. The cable between the AMPBOX and data acquisition

system can be over 100 meters.

Example:
You would like to connect a CMP10 pyranometer to the 4 to 20 mA inputs of your data acquisition system. The distance
between the two components is 45 meters.

Solution:
Order a CMP10, including adjusted AMPBOX and 50 meters of cable. When receiving the instrument and the cable with plug
attached, cut off 45 meters of cable and use the plug with 5 meters of cable to connect the instrument to the AMPBOX. The other 45
meters can be used to connect the AMPBOX to the data acquisition system.

Another option would be to purchase a separate cable with the specific length.

3.1 Installing AMPBOX on 4 to 20 mA current loop input

The next illustration shows how to connect the AMPBOX to a data acquisition system that can be configured to provide the power
for the current loop.

Note

Install the AMPBOX as close as possible to the radiometer. The cable between the AMPBOX and data acquisition

system can be over 100 meters.

E = irradiance level
I

out

= current recorded by the data acquisition

(4 to 20 mA)

I

offset

= zero offset from AMPBOX

(4 mA)

Gain setting =

(see chapter 2.2)

E

offset

= zero offset from irradiance

Examples of irradiance calculations:
• Standard gain setting of 2 mV / 1 mA
E = (I

out

- I

offset

)

* gain setting / S

instrument

• Adjusted gain for radiometers

(except pyrgeometers)

E = (I

out

- I

offset

)

* gain setting

• Adjusted gain for pyrgeometers
E = (I

out

- I

offset

)

* gain setting - E

offset

13

3. Installing the AMPBOX

+

GND

Datalogger

Radiometer

Active Current Loop

+

+

Example when using standard gain setting of 2 mV / 1 mA and output is 12,04 mA and sensitivity is 13,15 µV
E = (I

out

- I

offset

)

* gain setting / S

instrument

E = (12,04 mA - 4 mA) * 2 mV / 13,15 µV
E = 1222,81 W/m²

Example when using adjusted gain setting of 100 W/m² / 1 mA and output is 12,04 mA
E = (I

out

- I

offset

)

* gain setting

E = (12,04 mA - 4mA) * 100 W/m²
E = 804 W/m²

Example when using adjusted gain setting of 25 W/m² / 1 mA for the pyrgeometer starting at -300 W/m² and output is 16 mA
E = (I

out

- I

offset

)

* gain setting - E

offset

E = (16 mA - 4mA) * 25 W/m² - 300 W/m²
E = 0 W/m²

Note

Gain setting can be calculated by dividing the entire span by 16.

Example: -600 to +200 W/m² span = 800/16 = 50 W/m² per 1 mA

3.2. Installing AMPBOX on 4 to 20 mA input that does not supply the 4 to 20 mA current loop

Some data acquisition systems cannot supply the power for the 4 to 20 mA current loop. In such situations an external power
supply unit is necessary. This power supply will then supply the active current loop required for the AMPBOX.

Note

There is an internal diode that protects the AMPBOX from any damage when the power supply unit is connected

the wrong way around.

Note

The power supply should be suitable to supply 7.2 VDC for the AMPBOX + the voltage drop over the cable loop at

20 mA. The practical minimum is 7.5 VDC and the maximum is 35 VDC. The voltage drop over the 4 to 20 mA current

loop cables can be calculated when the specifications of the cable are known.

Note

Install the AMPBOX as close as possible to the radiometer. The cable between the AMPBOX and data acquisition system

can be over 100 meters. It’s therefore advised to install the PSU as close as possible to the data acquisition system.

Advertising