Konig Electronic Upper arm blood pressure monitor User Manual

Page 4

Advertising
background image

4

3.4 Storing your Measurement Results
To store your measurement results, you need to enter a filename in the blank space on the right hand side
of the screen and press “Save File”. The measurement results will be saved on your computer and is
displayed in graphics.

3.5 Averaging your Measurement Results
To know your average measurement results, select the measurement records you want to calculate and
click the AVG button. It will start to calculate the mean value automatically.

3.6 Blood Pressure Knowledge Guide
This instruction guide provides information about blood pressure and related health knowledge. Users are
able to study this information and use it to learn more about their health.

3.7 Description of Product Display

1. Pressure

Waveform

2. Integrated Pressure Waveform
3. Average

Measurement

Results

4. Blood Pressure Knowledge
5. Systolic

Pressure

6. Diastolic

Pressure

7. Mean Arterial Blood Pressure.

(Diastolic Blood Pressure + 1/3(Systolic Blood Pressure – Diastolic Blood Pressure))

8. Pulse Pressure. (20 mmHg < Normal Pulse Pressure < 60 mmHg)
9. Pulse

Rate

10. Start Button: Start the Measurement
11. Stop Button: Compulsively Stop the Measurement
12. Exit Button: Leave the Measurement Program
13. Save File Button: Save measurement results to PC in jpeg file format
14. Filename Column: Key in a file name
15. Link Indicator Light

4. Product Specifications

Name

Arm Type USB Blood Pressure Monitor

Model No.

HC-BLDPRESS250

Main Unit: 158 (L) × 50 (W) × 46 (H) mm

Product Dimensions

Arm Cuff: 470 (L) × 140 (W) mm

Measurable

Circumference of Wrist

Approx. 22 ~ 32 cm

Unit Weight

Approx. 150g

Pressure: 0~300 mmHg

Measuring Interval: 1 mmHg

Measurement Range

Pulse Rate: 30~200/min.

Accuracy

Pressure: within ±4 mmHg of reading

Pulse Rate: within ±5% of reading

Measurement

Oscillometric Method

Advertising