Labconco Biological Safety Cabinets 30468 Series User Manual
Page 10
Chapter 2: Prerequisites
Product Service 1-800-522-7658
4
Original instructions
If you intend to connect the biosafety cabinet to
an exhaust system:
 
NOTE: Only connect the cabinet to a suitable exhaust 
system, one dedicated to the cabinet itself, or 
dedicated to exhausting laboratory ventilation 
equipment. DO NOT connect the unit to the building’s 
HVAC system room exhaust. 
 
Examine the location to ensure that it accommodates 
the cabinet’s exhaust duct. The area directly above the 
cabinet’s exhaust port should be clear of structural 
elements, water and utility lines, or other fixed 
obstructions. There should be enough clearance to 
accommodate a 10-inch diameter duct. 
 
 
 
 
Avoid cabinet locations that require an elbow 
directly above the cabinet’s exhaust connection or 
an excessive number of elbows in the exhaust 
system. There should be a straight length 10 duct 
diameters long between the cabinet and any elbow, 
and between subsequent elbows. 
 
 
 
 
 
 
 
 
 
The Inlet Relief Valve located on the top of the cabinet is 
designed to draw a maximum of 100 CFM (2.83 m
3
/m).
Attempting to draw additional room air through the valve 
(room air exhaust), can result in unstable cabinet operation. 
 
 
 
 
 
 
 
 
 
Figure 2-1b
Figure 2-1c
Figure 2-1d