Chapter 2: Installation

Product Service 1-800-522-7658

6

The Two and Six-Place Kjeldahl Distillation Systems do not include a
surrounding safety enclosure. Labconco advises installing these products in a
fume hood, or if installed on the open laboratory bench, a safety shield should be
placed in front of the burners. High heat and boiling Sodium Hydroxide present a
safety hazard to personnel in close proximately to the Kjeldahl Distillation
Systems. Labconco also recommends spot exhaust ventilation for an open bench
installation to remove excess heat produced by the heaters.

Electrical

The Two-Place Kjeldahl Distillation System comes with an 8-foot, 3 wire cord
and plug and can be connected to a 115 volt 60Hz 11 Amp circuit.

The Six-Place Distillation Kjeldahl System has been wired at the factory per the
product model number ordered by the customer. Labconco model numbers ending
in 01 are 115 volts. Model numbers ending in 02 are 230 volts. Models ending in
03 are 230 volt, three phase. Amperage requirements vary depending on the
number of heaters per the Distillation rack. The voltage and amperage table below
provides the electrical requirements for the Six-Place Distillation Systems. A
qualified electrician will need to connect the facility electrical supply to the main
breaker box of the Digestion System.

1. Remove circuit breaker box cover panel.
2. Main line lead connection terminals are identified and connections

must be made accordingly.

3. Line leads to the apparatus must conform to local electrical codes.
4. Provide an electrical ground to the apparatus per code.
5. Before applying power to apparatus, check the electrical panel and

breakers for loose connections.

6. Reinstall box cover panel.
7. Power the breaker box and reset breakers to check circuits in the

apparatus.

8. All electrical wiring and connections must conform to local codes and

should be performed by qualified electricians. An earth ground must
be provided.


Six-Place Distillation System Electrical Code

Code #

AC Electrical Codes

01

115 volt, single phase, 50/60 Hz, 32 amps

02

230 volt, single phase, 50/60 Hz, 16 amps

03

208/230 volt, three phase, 50/60 Hz, 9 amps