14

Symptom

Probable Cause

Corrective Measure

Main pilot light does not
illuminate.

ON-OFF switch in off position

Turn switch to ON position

Main power fuse blown or
circuit breaker tripped

Replace fuse or reset circuit
breaker

25 Amp fuse on control unit
blown (8 Amp on Rapid
Digestor 4-Place)

Check continuity of fuse with
ohmmeter. Replace fuse if
blown. If fuse continues to
blow, locate cause.

ON-OFF switch defective

Replace switch. Refer to
Service Section for
replacement procedure.

No heat applied to unit.
Main pilot light
illuminated

Sensor defective

Check continuity of sensor
by placing ohmmeter across
pins 1 & 2 of female sensor
receptacle. If resistance
measurements show sensor to
be open, replace sensor by
following Service Section on
Sensor Replacement.

Indicator lamp not
illuminated.

Bulb defective

Check continuity of bulb
with ohmmeter and replace if
defective following Service
Section-Indicator Lamp
Replacement.

Defective resistor

Check continuity of resistor
for the row of heaters behind
indicator lamp not
illuminated using ohmmeter.

Defective ceramic heater

Check continuity of each
heater contained in the row
behind indicator lamp that is
not illuminated. Replace
defective heater following
Ceramic Heater Replacement
section.

Uneven heat applied to
unit (not applicable to
Rapid Digestor 4-Place

Strip heater defective

Check continuity of each
strip heater and replace
defective heater following
Strip Heater Replacement
section.

Advertising