Chapter 7: Troubleshooting

Product Service 1-800-522-7658

33

PROBLEM

CAUSE

CORRECTIVE ACTION

No vacuum/poor vacuum
(cont.)

Ice formed on Cold
Trap lid sealing surface

Defrost and wipe dry.

Secondary Trap
Cannister not fitted
properly

Tighten all connections to and from the
Secondary Trap Cannister.

Secondary Trap Insert
is spent

Replace with new insert.

New Secondary Trap
has moisture in it

Run vacuum pump for 24 hours to
remove moisture.

Recovery of condensate in
Cold Trap is less than
normally expected

Cold Trap is not ON

Check to make sure switch is ON and
condensing unit fan is moving air out
of the rear of the cabinet.

Cold Trap does not cool
down

Turn vacuum pump OFF and allow
Cold Trap to cool for at least 30 min. to
reach temperature.

-50°C temperature can be checked with
a solvent thermometer or digital
thermometer.

Frequent oil change needed
in pump

Secondary Trap insert
is spent

Change insert often.

Cold Trap is not
emptied after each run
and dried

Empty the traps (glass or stainless
steel) after each run and replace.

Vacuum too strong for
chemical

Use a Secondary Trap insert and
diaphragm pump.

Unit starts and shuts off

Latch optical sensor
improperly calibrated

Unplug power cord. Wait 10 seconds.
Plug in power cord.