Counting procedures

The scale must be in ON0 or ON2 modes.The display must be

showed “0.0”,

if not, press [ ]

key to set the display to “0”.

ON0 Mode:

1. Turn on the scale,"ON0" will show in display area, press [ ] key and all

display segments appears on LCD, and finally "0" will appear.

2. Place a given number of samples on the platform (the Sample Size should

be10、20、50 or 100 pieces).

3.Press [ ] key several times to put the scale in PCS mode. Press [PCS],

the display will show "xx pcs", press [PCS] to select XX value (10, 20,

50 ,100)

4. Press [ ]

key, the display will show “PASS” then “xx pcs”, then the scale

will recall the sample size selected and show the starting sample size on

the display,sample setting is done. Press [ ] key to set the display to
“0”.

5. Now you can put objects on the platform for counting pcs.

6.

Press [ ] key to return to weighing mode


ON2 Mode: (Precondition: Weight value of empty container need be

stored first through step ON1 of page 1)

1. Turn on the scale,"ON0" will show in display area, press [ ] key to

choose “ON2”,Then press [ ] key, all display segments appear on LCD
for a few seconds, and

finally it will show “0.0”.

2. Place a given number of samples into the container

(the Sample Size

should be10、20、50 or 100 pieces).

3.Press [ ] key several times to put the scale in PCS mode. Press [PCS],

the display will show P=XX,press [PCS] to select XX value (10, 20, 50,100)

4. Bottom up the scale on a stable and flat place, wait till "beep" heard. Put

the scale straight on a flat place, "P=xx" will show on display

5. Press the [ ]

key, the display will show “PASS” then “XX pcs”, then the

scale will recall the sample size you selected and show the starting sample

size on the display,sample setting is done. Press [ ] key to set the

display to “0”.

6. Now you can use count function by adding objects into container.

-3-

FUNCTION SETTINGS

1.Auto Shut-off

When the scale is in off mode, press and hold [PCS], then press [ ] to

turn on the scale. The display will show A-ON or A-OFF, press [PCS] key

to select A-ON or A-OFF. A-ON means auto shut off function active, A-OFF

means auto shut off function inactive.

2.Auto backlight

When the Scale in function setting mode, press [ ] key, the display will

show L-ON or L-OFF, press [PCS] to select L-ON or L-OFF. L-ON means

auto backlight function active, and L-OFF means auto backlight function

inactive.

3.Now function setting is finished. Press [ ] key to turn the power off .

CALIBRATION

Note: Calibration is done in factory. Don't re-calibrate the Scale,
unless it is not accurate.

How to calibrate
1. Before entering calibration mode, press [ ] to turn the power on for 1

minute.

2. Press and hold [ ] for 3 seconds to turn off the scale. Press and hold

[ ] , then Press [ ] to turn on the scale,

the display will show “CAL”

then the A/D value.

3. Press [ ], the display will show"-------" then flashed calibration weight

value(200g).

4. Place the calibration weight on platform.
5. Press [ ], the display will show "-------"then the A/D value

6. Now calibration is complete. Remove the weights.

7. Press [ ] key to turn the power off . Press [ ] again to turn on the