LW Measurements LWC 800 User Manual

Page 21

Advertising
background image

21


- Press HOLD key again, display will show Ani. Fd, press UNIT key to enter,
press ZERO key to select filter range between 0.5, 3, 5 and 10, press HOLD
key to confirm.
*** This is the filter range, 0.5d, 3d, 5d or 10d that the scale displays, for
example, if we use LWC 800lb x 0.2lb for weighing a 300lb person, if we set it
to be 0.5d, then the scale will show 299.8lb, 300lb, 300.2lb..., if we set it to be
3d, then the scale will show 299.4lb, 300lb, 300.6lb... 0.5d can get a much
more accurate weighing, but it won

’t be as stable; 10d will not obtain such an

accurate weighing result, but it will be much more stable.

- Press HOLD key again, the display will show Ani. Ft, press UNIT key to
enter, press ZERO key to select filter time between 1, 2, 3 and 4, press HOLD
key to confirm.
*** This is the filter time, 1s, 2s, 3s or 4s that will take to display the weighing.
Weighing will be more accurate if the user chooses 4 second (longer
displaying time), but not so accurate if 1 second (faster displaying time) is
chosen.

Press ON/OFF key to turn the scale off.

4.2.4 Hold function

- Press ON/OFF key to turn on the scale, place the item on the platform.

- Wait for the readings stable, press HOLD key, the

.

“ indicator will show


- Remove the item from the platform, the readings will still remain on the

display


- Press HOLD key to quit hold function mode