Martin Christ Delta 2-24 LSCplus User Manual

Page 93

Advertising
background image

Delta 1-24 LSC&

Delta 2-24 LSC&

10 Technical data

Version 04/2013, Rev. 1.2 of 20/02/2014 • sb

93

Translation of the original operating manual

Connection requirements
(without vacuum pump and

accessories)

Delta 1-24 LSC

plus

Delta 2-24 LSC

plus

Electrical connection:

3 x 400 V / 50 Hz

3 x 400 V / 50 Hz

Protection class:

I

I

IP protection category according to
DIN 60529:


11


11

Nominal power:

0.9 kVA

1.4 kVA

Nominal current:

3.5 A

3.5 A

Mains fuse:

10 A F

12 A F

Vacuum pump connection:

Depending on the type of vacuum

pump

Depending on the type of vacuum

pump

Filling quantity:

Delta 1-24 LSC

plus

Delta 2-24 LSC

plus

Refrigerant

See name plate

See name plate

Physical data (without vacuum
pump and accessories)

Delta 1-24 LSC

plus

Delta 2-24 LSC

plus

Dimensions
- Height:
- Width:
- Depth:


1,050 mm
860 mm
650 mm

(incl. vacuum flange connection)


1,050 mm
860 mm
650 mm

(incl. vacuum flange connection)

Weight:

approx. 215 kg

approx. 250 kg

Noise level according to DIN

45635:

__ dB(A)

__ dB(A)

EMC according to EN 55011:

Class B

Class B

Heat emission:

0.9 kW min.

1.9 kW max.

1.7 kW min.

3.0 kW max.

Equipment connections:

Delta 1-24 LSC

plus

Delta 2-24 LSC

plus

Vacuum connection:

Small flange connection DN25KF
(ISO 28403, DIN 2861)

Small flange connection DN25KF
(ISO 28403, DIN 2861)

Condensate drain valve:

Hose nozzle DN10

(outside diameter 12 mm)

Hose nozzle DN10

(outside diameter 12 mm)

Aeration valve:

Hose nozzle DN4
(outside diameter 8 mm max.)

Hose nozzle DN4
(outside diameter 10 mm max.)

Vacuum sensor:

SUB D-9

VSP 62

SUB D-9

VSP 62

Data interface (LAN):

RJ 45

RJ 45

Advertising