Figure 2-5, Open loop configuration with 1552 — 18, Figure 2-6 – Parr Instrument 6300 User Manual
Page 18: Open loop configuration — 18, Nstallation, 6300 c
Advertising

2
I
NSTALLATION
P a r r I n s t r u m e n t C o m p a n y
18
6300 C
ALORIMETER
E
XTERNAL
P
LUMBING
(C
ONTINUED
)
Figure 2-5
Open Loop Confi guration with 1552
O2 Regulator
DI Water
~Room Temp
Rinse Collection
1552
Water
Cooler
Tap Water
· <25 °C
· <85 ppm
· <60 psig
Drain
Figure 2-6
Open Loop Confi guration
O2 Regulator
DI Water
~Room Temp
Rinse Collection
Tap Water
· <25 °C
· <85 ppm
· <60 psig
Drain
5-10 °C
Line 1 & 2 – Maximum length of 10 feet, 1/4” OD, Polyurethane (Part Number HJ0025TB035)
Line 3 - Maximum length of 25 feet, 1/8” OD, Nylon (Part Number HX0012TB024)