5000 Series Multiple Reactor System
INSTALLATION
General Setup Guidelines
The Series 5000 Multiple Reactor
System requires at least 10 square feet
of work space on a sturdy bench or
table in a well ventilated area with
convenient access to an electric outlet.
If the tabletop is not heat resistant it
would be ideal to provide an insulated
pad on which to set the vessels when
hot.
Set the controller near the reactor on a
sturdy bench or table where there is
convenient access to an electrical outlet
capable of carrying up to 20 amperes.
Leave a space of at least twelve inches
between the controller and the heater
block so that the controller will not be
affected by radiant heat.
Installation
The system should be assembled
according to Diagram 1. Please note
that this figure illustrates the rear
view of all components in order to
clearly show the external wiring
details. From the front, the stations are
numbered 1 through 6, from left to right.