Physitemp Instruments BAT-12 User Manual

Page 4

Advertising
background image

3.0 OPERATING INSTRUCTIONS
3.1

The following instructions are found on the metal label fixed to the BAT-12.

They are reprinted here in case the label ever becomes defaced.

3.2

The special indications mentioned in 3.1 alert the user to unusual conditions

and can prevent the recording of erroneous readings.

3.3

When the instrument is switched on and a probe is connected, it should read

room temperature. This is normally about 25°C. Hold the tip of the probe between
the thumb and forefinger; if reading increases to 28-34°C, which is normal skin
temperature, it may be assumed that the instrument is working correctly. If the
instrument reads low, your hand may be cold. BAT-12 reads to 1/10°C.

3.4

Normal use of the thermometer is very simple. Switch ON, apply probe, read

temperature.

4.0 BATTERY REPLACEMENT
4.1

When a colon (:) appears in the center of the ditigal display, the battery needs

to be replaced. However, BAT-12 can be used without significant calibration error
for about one hour after the appearance of the colon.

4

TO MEASURE

TEMPERATURE

TEMPERATURE

RANGE

SPECIAL

INDICATIONS

BATTERY

REPLACEMENT

RECORDER

OUTPUT

Insert the plug of any Type T thermocouple
probe into the blue socket marked INPUT.
Switch ON. Apply probe. Read temperature.

Model BAT-12 reads -100° to +200°C
0.1° resolution

Probe is above max. temp. See range
above.

Probe is disconnected or broken

Colon between center digits indicates
low battery.

Switch OFF. Use coin to undo 2 screws in back
and slide case off. Remove battery from clip
and replace with similar 9V battery.

3.5 mm miniature jack socket on front panel.
10mV/°C

MODEL: BAT-12 SERIAL NO.

BAT-12

OPERATING NOTES

PHYSITEMP INSTRUMENTS INC
154 HURON AVENUE, CLIFTON NJ 07013 U.S.A. MADE IN

U.S.A.

0

18:88

1