Physitemp Instruments BFS-MP 6310 Rev1 Pak400 User Manual

Page 8

Advertising
background image

6.0

SPECIFICATIONS

6.1

CONTROLLER
Operating Range:

BFS-3MP and 5MP:

-35°C to ambient

BFS-30MP and 40MP:

-40°C to ambient

Digital Readout Resolution:

0.1°C

Accuracy:

±0.2°C

Ambient Operating Range:

15° - 45°C

AC Supply Operating Range: Voltage

100 - 240VAC 350VA

Input Current

3.2A max.

Frequency 47-63Hz

Power Output to stage:

BFS-40MP - 15VDC @ 10Amps
BFS-30MP - 12VDC @ 10Amps
BFS-3 & 5 MP - 5VDC @ 10Amps

Display:

Dual LED display of set and run Temperature

Safety Features:

Power shutoff during over-temperature condition caused by
inadequate coolant flow (50°C). Short circuit protection.

Size:

9 1/2” wide x 10” deep x 4 1/2” high

Weight:

5 lbs

Other Features:

Self-sealing water connectors.
Auxiliary switched AC output for ECV-1 or PTU-3. 2A max

6.2

STAGE
Temperature Range:

-40°C to ambient

Controlled Surface Area:

BFS-3MP

- 3 x 4cm

BFS-5MP

- 3.8 x 4cm

BFS-30MP

- 7.5 x 8cm

BFS-40MP - 7.1 x 8.3cm

Weight:

15ozs, including leads

Lead Length:

5ft

8