Physitemp Instruments TS4 User Manual

Page 4

Advertising
background image

2.3

ATTACH AC LINE CORD

This module is on the back panel, bottom left. There are four settings; 100V for unusually low
power situations, 120V, 220V and 240V. Confirm the selector is set for your requirements.

Slide clear plastic panel to left. If correct voltage is not visible on card, card must be pulled out
and replaced so that correct voltage is visible.

To change the position of the card, eject fuse by pulling on the FUSE PULL lever. Pull card for-
ward, rotate and reinsert.

Replace fuse and slide plastic cover back to the right. Attach line cord to controller.

2.4

ATTACHING WATER TUBES

Pump and Tank Unit is supplied with two water tubes.
All connections are self sealing when disconnected, to
prevent water spills. To connect, push in firmly. Squeeze
to disconnect.

Connect one of the tubes from outlet (OUT) of pump to
water inlet (IN) on rear of controller. Connect the second
tube from inlet (RETURN) of pump to outlet (OUT) on
the rear of the controller.

2.5 CONNECTING

THE

STAGE

Connect the two water tubes from the stage to the WATER connections on the front panel of the
controller. These tubes are interchangeable - direction of flow is not important.

Connect the black, corrugated cable to the polarized POWER OUTPUT connector on the front
panel. Twist locking nut to lock connector to controller.

Insert blue thermocouple connector into the socket labelled SENSOR B - STAGE.

2.6

FITTING THE STAGE TO YOUR MICROSCOPE

TS-4, TS-4ER

The black plastic base of the TS-4 Stage is the same size as a standard 1" x 3" microscope slide,
and fits into the mechanical stage. The stainless steel base slips under the mechanical stage
clamps, retaining the unit securely.

The upright heat exchanger section is positioned either towards or away from the operator to clear
the movements of the microscope objective turret.

4

BEFORE CONNECTING CONTROLLER TO AC POWER SUPPLY, CHECK THE SETTING

OF THE VOLTAGE REGULATOR MODULE