Position the gage relatively close to zero, then use the fine adjustment

screw to precisely position to zero. After final adjustment, replace the

safety cap.
Making a Measurement

To measure, first retract the spindle, then lower it gently to the work-

piece. The measured value or the deviation from the nominal value

(preset zero value) can be read directly on the dial.
An optional spindle lift cable (PT15053) is available. To attach, remove

the screw on the left of the case, then attach the cable release.
Changing Contact Points

The 647 contact point has a #4-48 NF thread. The 647M contact point

of model has a M2.5 thread. Be sure that your replacement contact

point has the same thread as the one being replaced.
To remove the contact point, press the spindle all of the way to the

upper stop and unscrew the contact point in a counter-clockwise

direction.
To install a contact point, press the spindle all of the way to the upper

stop. Then turn the contact point fully clockwise until it is firmly seated

to ensure accuracy.
Inspection

Inspect regularly to ensure proper operation and calibrate as needed

based on your frequency of use and accuracy requirements.