Test Products International 712 User Manual

Page 20

Advertising
background image

6.2 Function 2: - Manometer

Press the Func Key repeatedly to access the thermometer function. In this
mode the 712 functions like a differential manometer. The pump will stop
running when in this function.
1. Ensure you have

Pressure Sampling Tube connected to one or both of the Pressure Ports and
there are no restriction in the tubing (see figure below)

2. Zero the display by pressing the Down Arrow Key.
3. Connect the tube(s) to the device under test and read the pressure on the
display.

Other Features:

• Pressing the Up Arrow Key will scroll through mbar, kPa and inH2O
• Pressing the Down Arrow Key will Zero the Pressure reading
• Pressing the Scroll/Enter Key will toggle the ch2 temperature reading ON
and OFF

The 712 incorporates a differential manometer. This means when a single
hose is connected to the (+) port and pressure is applied, the 712 will read
positive pressure. If a single hose is connected to the (-) port and pressure is
applied, the 712 will read a negative pressure. If both ports are connected to
a pressure, the 712 will read the difference (P1 - P2) between the two.

6. FUNCTION SELECTION

Pressing the Func Key enables access to the different functions available on
the 712. Available functions are: Combustion Analyzer, Thermometer,
Manometer, and Date / Time display.

6.1 Function 1: - Thermometer

Press the Func Key repeatedly to access the thermometer function. In this
mode the 712 functions like a K-Type thermocouple thermometer. The pump
will stop running when in this function.

1. Ensure you have a 'K' type probe connected to one or both of the thermo-
couple sockets T1 or T2 (refer to figure below)
WARNING: - There is ONLY one correct way to connect the 'K' type thermo-
couple plug into the socket (see 4.2). Forcing the plug into the socket the
wrong way may result in damage to the instrument.

2. Touch the temperature probe to the item under test and read the displayed

temperature.

Other Features:

• Pressing the Up Arrow Key will toggle between ºC and ºF
• Pressing the Scroll/Enter Key will toggle the Differential Temperature ON
and OFF
• 'oPEn' will be displayed if no 'K' type probe is connected to the thermocou-
ple socket

The thermometer func-
tion allows you to per-
form temperature tests.

The pressure function
allows you to perform draft
and pressure tests.

Function Display 1 (Temperature)
T1 Temperature
T2 Temperature
Differential Temperature

Function Display 2 (Pressure)
Pressure Reading
T2 Temperature

Advertising