Accessories – Casella CEL AFC123 User Manual
Page 20
Advertising

Casella USA
17 Old Nashua Road #15
Amherst, NH 03031
Email
Web
www.casellausa.com
TF (800) 366-2966
AFC123
Personal
Sampler.
Page
20
Handbook No. HB3182-02 Issue2/JULY 1994
Accessories.
HK2
Eccentric to motor shaft
(Hexagon wrench)
HK304
Flow
adjustment
Service Hints:-
Symptom:- Pump does not run (green LED illuminated).
Possible cause:-
Service Hint
Battery fully discharged.
Replace with fully
charged battery.
Wiring loom disconnected at some point.
Locate and resolder.
Fault on joint.
Replace circuit
board.
Symptom:-Pump does not run (green LED illuminated).
Possible cause:-
Service Hint.
Eccentric bearing seizure.
Replace bearing.
Motor shaft seizure Replace motor.
Symptom:-Motor runs but pump inoperative.