Features

Backlit LCD display

Low battery indicator

Hold reading function

Fully waterproof

Fully guaranteed for 1 year

Auto off function

Double junction probe

(not replaceable)

Automatic Temperature

Compensation (ATC)

Successful calibration indicator

Selectable units for °C and °F

Power button / hold
Short press to turn on.
Short press to hold reading.
Long press to turn off.

Calibrate button
See calibration section.

Units button
Hold until units fl ash then

short press to change units.
Screen will change back

when no buttons have been

pressed for 3 seconds.

Quick guide

Storage

cap can be

placed onto

loop while

being used.

Low battery warning
Appears when

batteries are low.

Check mark/tick to

indicate successful

calibration
Disappears 30 days after

calibrating probe as a

calibration due reminder.

ATTENTION

If it dries, it dies!

Storage cap
The probe must not be allowed to dry out. Always place storage cap back onto

probe tip after each use. Add 3-5 drops of KCl storage solution to the round hole

(wetting seal) in the cap each week to keep the probe wet. See section 3.0

Keep your probe tip wet

at all times to avoid permanent damage