Calibration 5, Error messages, 0 error messages – Bluelab ppm Pen User Manual

Page 5: 0 calibration, English

Advertising
background image

5.0 Error messages

The following error messages appear for the following reasons.

4.0 Calibration

Calibration of conductivity is not required for this unit as it is factory calibrated.
However; if you wish to calibrate the unit follow the instructions below.

YOU MUST CLEAN THE PROBE BEFORE

CALIBRATING.
See section 2.0.

Rinse probe in fresh water and place it in a known

standard solution. See chart below for the correct

solution.
Wait for reading to stabilize.

Hold down the cal button for 3 seconds until CAL

appears.
Release button and CAL P should be displayed. If Err

is displayed check the probe is clean and that the

calibration solution is fresh and uncontaminated.

A check mark will appear on the screen to indicate that

the calibration was successful. The check mark will

disappear after 30 days. To reset back to factory default

remove/replace battery.

NOTE: If you need to test or calibrate in a 1500 ppm

solution, you MUST set the pen to EC, then multiply your

result by 540. If calibrating, multiply 2.8 by 540

(2.8 x 540 = 1512).
This unit DOES NOT measure in the 540 ppm scale.

Temperature

over range

ppm over range

Temperature error

Hardware error

Temperature

under range

English

5

1

2

3

4

EC

ppm 500

(TDS)

ppm 700

(EC x 700)

Solution value

2.77

1385

1940

Displayed value

2.8

1390

1940

1385

ppm

Hold for 3

seconds