Introduction 3, Preparing for use, 0 introduction to the bluelab ph meter – Bluelab pH Meter User Manual

Page 3: 0 preparing for use, English

Advertising
background image

English

3

1.0 Introduction to the Bluelab pH Meter

The Bluelab pH Meter has two press buttons; ‘calibrate’ and power. The power button

requires a short press; release in about one second. The ‘calibrate button requires a

long press; hold for at least three seconds and release when the display starts fl ashing.

Turning the pH meter on and off

A short press of the power button will turn the pH meter on. The pH meter

automatically turns off after approximately four minutes if no buttons are pressed.

If the pH meter turns off before the reading is taken, short press the power button to

turn the pH meter on again.

2.0 Preparing for use

The following tasks must be performed before the Bluelab pH Meter is used for the

fi rst time.

Insert batteries.
See section 5.0.

Connect pH probe
Connect the pH probe to the pH meter by lining

up the lugs of the BNC fi ttings. Fasten securely

by pushing the pH probe connector on and

twisting one quarter turn.

Remove the storage cap
Remove the pH probe storage cap by gripping

the top of the cap and gently twisting the base

one rotation clockwise to loosen slightly. Next

slowly slide the cap off the pH probe. DO NOT

completely remove the base of the cap from the

top of the cap.

CAUTION: When the pH probe is not in use, add

enough Bluelab pH Probe KCl Storage Solution to

the storage cap so the probe tip is covered. Then

replace the cap and store in a secure place.

DO NOT use RO (Reverse Osmosis), Distilled

or De-ionized water. Pure water changes

the chemistry in the reference, causing the

probe to die.

Calibrate the pH
Calibrate the pH meter by following the

instructions in section 3.0 of this manual.
This must be done before the

pH Meter is used for the fi rst time.

1

1

2

3

4

Attaching the

Bluelab pH Probe to the Meter

Removing pH probe

storage cap

Ensure probe tip is covered by

the KCl storage solution in cap

Inserting

Twisting

Attached

See section 3.0
for calibration
steps