6 troubleshooting – Bronkhorst E-7000 User Manual

Page 66

Advertising
background image

BRONKHORST HIGH-TECH B.V.

page 66

9.17.004

6 TROUBLESHOOTING



•

LCD is not lit after 'power on'

•

check power input

•

check

fuse

•

Improper

reading

•

check module settings

•

No

reading/control

•

check instrument cabling

•

Improper system operation (alarm, counter)

•

check module settings

•

Instrument does not respond to setpoint given at
keyboard

•

setpoint source at operation menu should enable
setpoint at "keyb. & FLOW-BUS". (other settings
means different behaviour)

•

setpoint could have been overruled at an alarm
situation or counter limit (first perform reset)

•

setpoint slope has a large value. (setpoint will
slowly change in time) Change slope value if
necessary.

•

setpoint could be changed by other devices
connected to the FLOW-BUS.

•

check polynomial factors for setpoint and
measure at instrument menu.

•

check communication with FLOW-BUS.
If applicable

•

Setpoint menu cannot be reached

•

check setting of connected instrument type at
instrument menu. For sensor types setpoint
facilities will be disabled.

•

Counter/alarm is not responding to reset

•

check if reset-source has been enabled at reset
menu of reset counter/alarm.

•

Instrument seems to give inaccurate values

•

check if correct fluid set (with calibration factors)
has been selected (see fluidname).

•

Display shows ‘---‘

•

check if the operating instrument is active

•

check if the operating instrument is still at the
same address.

•

check instrument and module cabling

•

Other

problems

•

check instrument and operating module
parameters.


Advertising