Flowserve X-100 User Manual

Page 5

Advertising
background image

5

of the gland may be necessary with

some pump designs to permit seal

piping to clear the bearing housing.

Refer to the assembly drawing

that is supplied with the seal for

proper positioning. Bolt the back

plate in place on the bearing

housing.
2.3 Assemble the pump, adjust the

bearings, set the impeller and

connect the coupling so

Install pump backplate

Figure 4

Tighten drive collar

set screws

Figure 5

that the shaft is in its operating

axial position. Any subsequent

adjustment of the shaft requires

resetting of the seal. Connect

pump piping, allow no pipe strain

on the pump casing.

2.4 Position the X-100 cartridge gland

with the gland gasket in place against

the seal chamber (stuffing box) face

and

tighten the nuts evenly, cross

staggering the adjustment of the nuts.

The gland nuts should be torqued to

a maximum of 13 N-m (10 ft-lbs).

Excessive gland nut pressure can

result in distortion of the stationary face.

2.5 Tighten the set screws on the X-100

cartridge drive collar with the Allen

wrench provided. See Figure 5.

2.6 Remove setting devices by removing

the cap screws with the Allen wrench

provided. See Figure 6. Save the setting

devices for use if pump impeller must be

reset or if the X-100 is removed for

maintenance.
2.7 Turn the shaft by hand to ensure free

operation.
2.8 Pipe up the gland connections to the

X-100 seal, see section 3.

2.9 See Operational Recommendations,

section 4, before starting pump.

Remove setting devices

Figure 6

Advertising