Manual Supplement

Impulse 6000D/7000DP Users

10

4/13

Change #3,

48038

On page 38, under Pacemaker Input, change:

From:

Accuracy …………..±1 % non-inductive (<2

H)

To:

Accuracy……………±2 % non-inductive (<2

H)

Change #4

On page 3 add the following Caution:

Caution

To avoid damage to the product or defibrillator under test, do not use adhesive pads on the Analyzer. For
defibrillators with metal paddles, use Fluke Biomedical electrode adapters.

Change #5,

63282, 65427

On page 2, delete CAT I from the Symbols table and add:

Conforms to relevant South Korean EMC Standards

.