Specifications – Fluke Biomedical 07-453 User Manual
Page 16

Section 6
Specifications
Model 07-453 SPECIFICATIONS
Accuracy
•
AC Input / X-ray Pulse - +/- 1 Count
•
DC Input - 2% +/- 1 Count
o
Measures at 70% of peak value
Sensitivity
•
AC Input - 90 VAC minimum
•
X-ray input - 50kVp, 5mA at 2.0 inches from top surface of case, pointed to
target on case.
Range
•
9999
Pulses
•
9999
Milliseconds
Display
•
0.3" Liquid Crystal
Power
•
9 Volt Battery
Low Battery Indicator
•
"Lo Batt" Appears in display when battery voltage reaches 5.3 Volts +/- 0.3 Volt.
Size
•
80 X 147 X 40 mm
•
3.15 X 5.8 X 1.6 inches
Battery Life
•
48 Hours minimum
o
With Duracell MN1604 Alkaline or equivalent
AC Input Jacks
•
130 Volts AC maximum
•
90 Volts AC minimum
•
Input circuit not affected by reversed polarity
Optional Accessories
•
Remote x-ray sensor
•
Carrying
case
•
AC test leads with tip plugs