Specifications – Fluke Biomedical 07-453 User Manual

Page 16

Advertising
background image

Section 6

Specifications

Model 07-453 SPECIFICATIONS

Accuracy

•

AC Input / X-ray Pulse - +/- 1 Count

•

DC Input - 2% +/- 1 Count

o

Measures at 70% of peak value


Sensitivity

•

AC Input - 90 VAC minimum

•

X-ray input - 50kVp, 5mA at 2.0 inches from top surface of case, pointed to

target on case.


Range

•

9999

Pulses

•

9999

Milliseconds


Display

•

0.3" Liquid Crystal


Power

•

9 Volt Battery


Low Battery Indicator

•

"Lo Batt" Appears in display when battery voltage reaches 5.3 Volts +/- 0.3 Volt.

Size

•

80 X 147 X 40 mm

•

3.15 X 5.8 X 1.6 inches


Battery Life

•

48 Hours minimum

o

With Duracell MN1604 Alkaline or equivalent


AC Input Jacks

•

130 Volts AC maximum

•

90 Volts AC minimum

•

Input circuit not affected by reversed polarity


Optional Accessories

•

Remote x-ray sensor

•

Carrying

case

•

AC test leads with tip plugs

Advertising