Fluke Biomedical 07-645 User Manual

Page 8

Advertising
background image

Nuclear Associates 07-645
Operators Manual

1-4

Table 1-1.

Percent contrast for Fluoro-Test Threshold Contrast Resolution Plates with 2

mm Cu Attenuator

kVp

Hole

Number

50 60 70 80 90 100 110 120 130

0.5 0.84

0.66

0.56

0.50

0.46 0.43 0.41 0.39 0.38

1 1.67

1.31

1.11

1.00

0.92 0.87 0.82 0.79 0.76

1.5

2.51

1.97

1.67

1.50

1.39

1.30

1.24

1.18

1.14

2

3.34

2.62

2.23

2.00

1.85

1.74

1.65

1.58

1.52

2.5

4.18

3.28

2.79

2.50

2.31

2.17

2.06

1.97

1.90

3

5.01

3.94

3.34

3.00

2.77

2.60

2.47

2.37

2.28

3.5

5.85

4.59

3.90

3.50

3.23

3.04

2.88

2.76

2.66

4

6.68

5.25

4.46

4.00

3.70

3.47

3.30

3.16

3.03

4.5

7.52

5.90

5.01

4.50

4.16

3.91

3.71

3.55

3.41

5

8.35

6.56

5.57

5.00

4.62

4.34

4.12

3.95

3.79

5.5

9.19

7.22

6.13

5.50

5.08

4.77

4.53

4.34

4.17

6

10.02

7.87

6.69

6.00

5.54

5.21

4.94

4.74

4.55

6.5

10.86

8.53

7.24

6.50

6.01

5.64

5.36

5.13

4.93

7

11.69

9.18

7.80

7.00

6.47

6.08

5.77

5.53

5.31

7.5

12.53

9.84

8.36

7.50

6.93

6.51

6.18

5.92

5.69

8

13.36

10.50

8.92

8.00

7.39

6.94

6.59

6.31

6.07

8.5 14.20

11.15

9.47

8.50

7.86 7.38 7.01 6.71 6.45

9 15.04

11.81

10.03

9.00

8.32 7.81 7.42 7.10 6.83