Fluke Biomedical 18-222 User Manual

Page 12

Advertising
background image

2-4

Table 2-2. Actual vs. Simulated Linear Attenuation Coefficients (u)

30% Glandular

100% Glandular

keV Actual Simulated keV Actual Simulated

10.0 3.400820

3.262850 10.0 4.919490

4.685870

15.0 1.112980

1.098010 15.0 1.560170

1.535910

20.0 0.574784

0.574784 20.0 0.768012

0.768012

30.0 0.302201

0.304501 30.0 0.368387

0.371520

40.0 0.232988

0.234800 40.0 0.268827

0.270928

50.0 0.205005

0.206445 50.0 0.229969

0.231386

60.0 0.189907

0.191118 60.0 0.209919

0.210931

80.0 0.172653

0.173673 80.0 0.188308

0.189033

100.0 0.161710

0.162638 100.0 0.175442

0.176050

50% Glandular

100% Adipose

keV Actual Simulated keV Actual Simulated

10.0 3.812000

3.622000 10.0 2.975010

2.837940

15.0 1.234110

1.213450 15.0 0.995056

0.960624

20.0 0.627163

0.627163 20.0 0.530186

0.511345

30.0 0.320188

0.323307 30.0 0.294058

0.280331

40.0 0.242758

0.245163 40.0 0.233268

0.220533

50.0 0.211829

0.213702 50.0 0.208080

0.195861

60.0 0.195389

0.196938 60.0 0.194093

0.182268

80.0 0.176952

0.178232 80.0 0.177546

0.166402

100.0 0.165485

0.166639 100.0 0.166690

0.156112

70% Glandular

keV Actual Simulated

10.0 4.231200

3.984430

15.0 1.357150

1.329930

20.0 0.679992

0.679992

30.0 0.337905

0.342249

40.0 0.252103

0.255583

50.0 0.218186

0.220987

60.0 0.200393

0.202773

80.0 0.180781

0.182797

100.0 0.168807

0.170643






Our simulated materials are formulated to maximize
simulation properties at 20 keV for the mammographic
energy range and 70 keV for the diagnostic energy range.



NOTE