Fluke Biomedical 18-252 User Manual

Page 4

Advertising
background image

Nuclear Associates 18-252
Operators Manual

1-2

The Contrast Detail Phantom for Mammography is easy to use. Simply place the phantom on the image
receptor surface in the same position as a breast. Position the x-ray tube and compression device as in a
craniocaudal examination. When using the phantom on prone position breast biopsy systems, use the
rotating top plate of the phantom and the compression device to secure the phantom against the image
receptor. Choose the appropriate kV and mAs factors (26 kV and 60 mAs works well on most systems),
or select automatic exposure control.

A good imaging system should resolve at least the following objects:

Row

Number

Minimum Number

of Objects

Detected

1 6

2 6

3 5

4 4

5 2

6 1

7 0

Minimum detectability score: 24/49

Advertising