General Information

Procedure

1

1-1

Section 1

General Information


1.1 Procedure

1. Place an appropriately sized cassette loaded with film in the normal orientation in the image receptor

holder.

2. Load film in the large cassette with the emulsion side of the film away from the intensifying screen.

3. Place the large cassette on top of the image receptor holder with the back of the cassette toward the

x-ray source, and ensure that the large cassette extends beyond the image receptor holder on the
chest-wall side by about 1 cm.

4. Place the collimator to be evaluated in position.

5. Remove the compression paddle. (The compression paddle should be removed before placement of

the collimation template to ensure sharp demarcation at the edges of the light field.)

6. Turn on the collimator light and align the rulers so that the “O” mark is on the edge of the light field on

all four sides. Position the AEC detector so the metal ruler at the chest wall is not covering it.

7. Replace the compression paddle. Position it so it just rests on the top of the peg holding the ruler for

the compression paddle edge alignment. This will set the height of the compression paddle 4.2 cm
above the image receptor. (If using thin cassettes use the 2.2 cm peg, if using the thicker cassettes
use the 1.7 cm peg.)

8. Place the right angle edge of the metal ruler snugly against the outside edge of the compression

paddle.

9. If necessary, to achieve and adequate exposure place a sheet of acrylic or BR-12 attenuating

material on top of the paddle, so that all radiation reaching the cassettes must pass through the
attenuator. Make an exposure using AEC.

10. Repeat steps 1 through 9 for all routinely used collimator/bucky/compression paddle combinations

and target materials. When testing the large image receptor, the top cassette may be positioned
diagonally to capture all four edges of the x-ray field, or two large cassettes may be used on top of
the image receptor holder. (The collimator test may also be done using a non-screen cassette on top
of the breast support, no attenuator, and a manual technique of approximately 26 kVp and 12 mAs.)
When measuring collimation in the magnification or spot compression mode, the peg used for the
compression paddle edge should be moved to the inner position of the template.

1.2 Date Analysis and Interpretation

When analyzing your image, follow the guidelines in the Medical Physicists section of the ACR manual.
The only difference is that you will not need to measure anything. You can simply view your image and
record the deviation.