Introduction, 1 general description, 2 application – Fluke Biomedical 6000-530B User Manual

Page 5: 3 specifications

Advertising
background image

Introduction

General Description

1

1-1

Section 1

Introduction

1.1 General Description

The Model 6000-530B Image Intensifier Ion Chamber is a low profile ion chamber designed to measure
diagnostic
x-rays.

1.2 Application

The Model 6000-530B Image Intensifier Ion Chamber is specifically designed for measurement of
exposure rate at the input phosphor of fluoroscopic image intensifier tubes. Its sensitive area and form
factors have been adapted for placement inside the spot film tray of common image intensifier systems.
Also included is a detachable handle for easy insertion and removal. When the host instrument is a Model
6000 series or a Model 4000M+, precise measurements may be made at low exposure rates (down to 10
μR/min) if the Model 6000-530B Preamplifier is used (See the Model 06-524- 2000 manual for operating
instructions).

1.3 Specifications

Chamber

Volume

150 cm

3

Nominal Sensitivity

45 nC/R

Housing Material

Acrylic

Housing Color

Clear

Housing Size

6.5 in diameter (16.5 cm diameter)

Surface Area

110 cm

2

, 5.5 in diameter

Weight

30 oz (840 g)

Handle Length

16.5 in (41.9 cm)

Height

0.720 in

Cable Length

10 ft (3 m)

Cable Termination

BNC/Banana

Window Thickness

.110 in (2.79 mm)

Total Attenuation

0.275 mm AI equivalent @ 30 keV

Operating Voltage

200 – 300 V DC

Compatible With

Model 6000 Series, 4000M+, & RAD-CHECK® PLUS