Figure 2. disinfection basin test setup – Fluke Biomedical ULT800 User Manual
Page 19
Advertising

Ultrasound Transducer Leakage Current Tester
Using the ULT800
9
ULT800
TEE TRANSDUCER
LEAKAGE CURRENT TESTER
INSTRUCTIONS:
1. Connect the transducer and the adapters to the ULT 800.
2. Press the ON/TEST button and wait for the READY light.
3. Select CONDUCTIVITY and press the ON/TEST Button-
Wait for the PASS light.
4. Select LEAKAGE and press the ON/TEST button again-
observe PASS or FAIL light. If test fails, DO NOT USE
TRANSDUCER. Refer to the user’s manual.
ON
READY
PASS
FAIL
LOW BAT
TEST
CONDUCTIVITY
LEAKAGE
Biomedical
Transducer-Specific
Adapter
Dual Conductivity
Electrodes
Fluke
ULT800
X
ecy02f.eps
Figure 2. Disinfection Basin Test Setup