Nuclear Associates 76-823,76-824,76-825
Operators Manual

1-2

Linearity Insert

Acrylic, 7-7/8" OD x 2" thick. Crossed grid of channels are cut out in a 1 ½” deep square pattern.






















Figure 1-1

Exploded View of Phantom

1.3 General Instructions

A. Remove the cover of the tank by unscrewing all the thumbscrews. Check the packing material

inside the phantom tank carefully for the plugs, o-ring and additional thumbscrews. Replace the
inserts as desired.

B. Place a small amount of silicone grease on a paper towel (s topcock or vacuum grease is

acceptable), and lightly coat the entire o-ring. Also lightly grease the o-ring groove in the tank, using
a cotton swab. Place the ring in the groove, being careful not to over-stretch the o-ring. The sloped
inner side of the groove will hold the o-ring in place.

C. Whenever the tank is opened, inspect the o-ring. Re-grease it when it appears dry (once or twice

per year).

Tank

"Cold"
Lesion
Insert

"Hot"

Lesion

Insert

Linearity/

Uniformity

Insert