3 troubleshooting – Fluke Biomedical 943-35 User Manual

Page 18

Advertising
background image

Victoreen 942-35,943-36,943-36H, 943-37,943-237A
Operators Manual

4-4

943-35, 943-36, 943-36H 200 CPM

943-37

400

CPM

943-237A

3000

CPM

3. Place

a

133

Ba button source (P/N 844-36-15), blank side up in the slide drawer of the standard

geometry, and insert drawer into geometry.

4. The expected count rate for the button source should be determined. It is recorded on the factory

calibration data sheet. Correction must be made for decay. (The half-life of the source is 10.74
years).

5. Run a 2-minute count (10,000 CPM minimum), using the

133

Ba source.

6. Adjust the high voltage power supply to the detector so that the net count rate corresponds to the

expected count rate calculated in step 4. (Net count rate is indicated count rate minus background
count rate). A ± 2% tolerance is acceptable.

7. The count rate will increase and decrease in conjunction with the high voltage setting. If the high

voltage reaches the maximum allowable setting without achieving the necessary count rate,
suspect a defect in the detector, wiring or readout device.

8. If additional button sources were provided, repeat step 5 for each source. A 6% tolerance is

acceptable.

9. If the above sources are not within ± 6%, the gain may be incorrectly set on the 356 keV

133

Ba

photopeak and not on the 80 keV

133

Ba photopeak. Using a

137

Cs source and an oscilloscope, verify

and adjust the high voltage so that the

137

Cs photopeak is at approximately 2.3 Volt average (1.65

Volts for Americium regulated detectors). Step 6 should then be repeated being sure to adjust the
high voltage in the negative direction only.

Calibration Without Transfer Button Sources

In the event secondary transfer sources are not available, the detector high voltage may be set to the
value stated on the calibration data sheet by using a

137

Cs source and an oscilloscope. The average

height of the

137

Cs photopeak is approximately 2.3 Volts, at the high voltage stated on the calibration data

sheet, and may be checked to determine proper detector alignment. Without a set of secondary sources,
however, the overall detector response in terms of counts per minute, μCi per cm

3

must be determined by

the user's primary calibration.

4.3 Troubleshooting

Extreme care must be used when troubleshooting a
system that has power applied. All standard
troubleshooting precautions apply.



Extreme care must be used when troubleshooting
in close proximity to the high voltage output.



Once a problem has been located, remove all
power before continuing with the repair.



WARNING