Installation, 1 mounting, 2 power – Fluke Biomedical 05-433 User Manual

Page 9: 3 optional power, 4 grounding

Advertising
background image

Installation

Mounting

3

3-1

Section 3

Installation


3.1 Mounting

The mounting bracket supplied may be attached to any convenient vertical mounting surface. Before
mounting the bracket to the wall, however, make sure that the six-foot converter power cord will reach the
nearest power outlet.

The wall-mounting bracket comes supplied with double-sided foam tape already attached. Just peel off
the protective paper covering and firmly apply to the mounting surface.



Once in place, the bracket cannot be moved or
removed without destroying the tape. Make sure of
your installation site before applying. If you wish to
attach the bracket with screws, mounting holes are
also provided.

3.2 Power

The plug on the end of the converter power cord should first be inserted into the jack on the bottom of the
Primalert. The converter should then be plugged into the wall outlet.

3.3 Optional Power

The Primapak Backup battery pack (P/N 05-441) provideds a sourced of 12 to 16 VDC power.



The converter plug supplies this voltage.


The plug is a Switchcraft S760, or equivalent, with the sleeve negative and pin positive. The current
drain at background radiation levels is about 25 mA and about 250 mA at full alarm levels.

3.4 Grounding

If it is desired to ground the instrument with a separate ground wire, as may be required under certain
codes, Figure 3-1 shows how to accomplish this. Simply remove the back cover by loosening the single
screw in the center of the cover. Note the position of the ground terminal in Figure 3-1. Attach the wire by
running it up through the small hole between the jacks. At least a #16 gauge wire should be used for this
purpose.

CAUTION

NOTE