General information, 1 product description, 2 specifications – Fluke Biomedical 05-582 User Manual

Page 5

Advertising
background image

General Information

Product Description

1

1-1

Section 1

General Information

1.1 Product Description

The PRIMA® 7 is a general-purpose survey meter designed for the measurement of gamma radiation. It
can be used to determine *TRANSPORT INDEX of packages containing gamma emitting radioactive
material. The telescoping rod, when fully extended, places the detector of the PRIMA 7 at a distance of
one meter from the package being monitored. This is the distance at which the TRANSPORT INDEX is
measured.

Some early models of the PRIMA 7 have a telescoping measuring rod that is three feet long. On these
models, the rod, when fully depressed, protrudes from the top of the case about ¼ inch. On models with
a one-meter rod, the rod, when fully depressed, will protrude about ¾ inch.

1.2 Specifications

Range

0-99 mR/h, in 1 mR increments with round-off

Display

2 digit L.E.D., 0.3 in. high

Indicators

Read indicator lights up after 4-second measurement cycle is complete

Display blanks over 99 mR/h

Low-battery indicator flashes when battery needs to be replaced

Count indicator flashes on or off for each detector pulse

Energy Response

+40%, -12% from 50 keV to 1.2 MeV

Accuracy

± 10% or ± 1 digit, whichever is greater Calibrated for

137

Cs

Detector

Compensated halogen-quenched GH tube

Control

Press to measure (4 seconds); hold to read

Response Time

4 seconds

Batteries

4 type AA alkaline cells; over 10,000 measurements

Operating Environment 10°F to 120°F (-12°C to 50°C) with alkaline cells

Dimensions (HxWxD)

6 in. x 3 ½ in. x 1 ½ in.

(15.2 cm x 8.9 cm x 3.8 cm)

Weight

12 oz (340.2 g)


* U.S. Department of Transportation, Title #49