Troubleshooting

Troubleshooting

3

3-1

WARNING

NOTE

Section 3

Troubleshooting

An electrical shock hazard exists between the high
voltage supply and ground.

3.1 Troubleshooting

Refer to the Table 3-1 for symptom/cause/corrective action.

Table 3-1.

Troubleshooting Table

Symptom

Probable Cause

Corrective Action

The words “No Detector” is
displayed on the Model 190.

Probe Adapter Module not
plugged in.
Probe Adapter Module female
plug not correctly aligned with
the pins in the 190.

Make Sure The Unit is Off
During The Corrective Action.
Either plug in the module or
remove, check the pins, and
carefully re-plug.
If required, straighten the pins
with a small screwdriver.

Refer to the Model 190 Manual for other symptoms,
error messages, probable causes and corrective
actions.

.